TT-APZ-02 Audibert-Arnu Dilatation Test System
I. Product Introduction:
The TT-APZ-02 series fully automatic Audibert-Arnu dilatation tester independently developed by our company is a special equipment for determining coal caking property. It fully complies with the technical requirements of the national standard GB/T5450-1997 'Test for Audibert-Arnu Dilatation of Bituminous Coal'. The software operating environment is Windows 10 system. It is available in two models: one for testing two coal samples simultaneously, and the other for testing four samples simultaneously. The instrument is easy to operate, safe and reliable, and is the newest instrument for Audibert-Arnu dilatation test of bituminous coal. It is an indispensable coal quality analysis instrument for inspecting coal quality parameters, coal development, production, import and export, and coal blending for coking.
II. Technical Features:
1. The TT-APZ-02 fully automatic Audibert-Arnu dilatation tester is easy to use with accurate temperature control. It has a high degree of automation, automatically calculates the b value of the Audibert-Arnu dilatation index of bituminous coal, and prints the test result report of the Audibert-Arnu dilatation index of bituminous coal.
2. Uniform heating furnace temperature, accurate data, microcomputer high-precision temperature control technology, computer data acquisition and storage.
3. Current and voltage display method: digital
4. Power regulation method: 4-20ma PID control
5. Data, heating curve, and volume curve generated during the test are stored and recorded after the test for future observation and comparison.
III. Technical Parameters:
1. Control program: 0℃ to preheating temperature heating rate of 5℃/min, preheating temperature to 600℃ heating rate of 3℃/min
2. Temperature control accuracy: between preheating temperature and 600℃, temperature difference < ±1℃, resolution 1℃
3. Displacement measurement range: 0-200mm, accuracy 0.5%
4. Temperature measurement element: K-type thermocouple
5. Expansion rod weight: 150±5g
6. Temperature field at 180mm upward from the bottom of the expansion tube, deviation < ±3℃
7. Temperature display error: < ±1℃
8. Rated power: 2KW, Rated voltage: AC220V 50HZ, Heating furnace power: 1.5KW
9. Heating accuracy: temperature rise of (15±1)℃ within 5 minutes
10. Working environment: ambient temperature 0℃ ≤ T ≤ 40℃, humidity ≤ 85%, no strong magnetic field, no severe vibration or corrosive gas around.
Contact: Manager Zhao
Mobile phone: 86 15642202798
Consultation hotline: 400-805-1057
Landline fax: 86 0412-5291088
E-mail: lnttznkj@163.com
Company Address: No. 26 Yongning Street, Yongfa Street, Tiexi District, Anshan City, Liaoning Province
Official website: www.lnttznkj.com
Scan to follow